PuriLab Analytical
COA/BAC-2026-004Verified
Save as PDF

Certificate of Analysis

PuriLab Analytical

Bacteriostatic Water for Injection

BAC-2026-004Lot NOT PROVIDEDAnalyzed July 20, 2026

Timeline

Ordered
July 20, 2026
Received
July 20, 2026
Analyzed
July 20, 2026

Sample

Name
Bacteriostatic Water for Injection
Expected
Benzyl Alcohol (preservative)
Amount
Aqueous solution, multi-dose vial

Testing

Client
Anonymous
Test type
Preservative identity & content determination
Method
HPLC-UV
Content, how much is in it
1.55%w/v
Label claim: 0.9% w/v (typical)
Identity, what it isMatch
Confirmed
Retention-time match vs. reference standard

Chromatogram

HPLC-UV

mAU00.511.522.533.544.55Retention time (min)2.481.30
2.48 min, Benzyl Alcohol
1.30 min, Unretained peak (void volume)

Analyst Comments

Clear, colorless aqueous solution. Preservative identity confirmed against a reference standard. Measured benzyl alcohol content is 1.55% w/v, above the 0.9% w/v typical of commercial bacteriostatic water, consistent with evaporative loss of water from an aged or repeatedly accessed vial. This level remains well below concentrations of toxicological concern. No additional peaks greater than 1% of total area were detected. Purity is not reported for this sample type.

Analyst Signature

Max Barrett
Lab Director, PuriLab Analytical
Analysis date: July 20, 2026

Verification Code

BAC-D4H8-M2VT
Verify at purilab.org/verify/BAC-D4H8-M2VT

Scan to Verify

BAC-D4H8-M2VT

Results apply solely to the sample as received and are not representative of any larger lot, batch, or production run. Findings are limited to the analytes stated above. PuriLab Analytical is an independent testing laboratory, is not CLIA certified or ISO 17025 accredited, does not sell or distribute any substance, and makes no representation as to the safety, legality, or fitness of any material for any use.